The commonest error in reading an ECG is stopping at the first abnormality. Before interpreting anything, check that the tracing is a complete, labelled 12-lead recording with a rhythm strip at the standard 25 mm/s and 10 mm/mV, and look for misplaced leads: a negative P wave and QRS complex in lead I suggests reversed arm electrodes or dextrocardia. The statement printed by the machine is a draft to be checked. Errors in computer interpretation persist despite better algorithms, and inexperienced readers may accept them without question [11].
| Step | What to look at | Where readers go wrong |
|---|---|---|
| Rate | Regular rhythm: 300 divided by the large squares between two R waves. Irregular: QRS complexes on the 10-second strip times 6 | Using the regular-rhythm method on an irregular rhythm |
| Rhythm | P waves, their link to each QRS, and regularity | Missing flutter with 2:1 block behind a steady rate of 150 |
| Axis | Leads I, II and aVF | Overlooking a new axis shift |
| Intervals | PR 120 to 200 ms; QRS of 120 ms or more is broad; QT corrected for rate | Missing a wide QRS from potassium or a sodium-channel blocker |
| Morphology | Q waves, R wave progression, hypertrophy, bundle branch block, delta waves | Reading voltage alone as hypertrophy |
| ST and T | Each territory, its reciprocal leads, and any earlier tracing | Accepting early repolarisation in a patient with chest pain |